Tonometry
Indications:
- to measure intraocular pressure – normal IOP -12 – 21 mmHg
- perform after visual acuity testing
Contraindications:
- suspected globe rupture or penetrating injury
- history of recurrent corneal erosions
- severe bullous keratopathy
- corneal abrasions
- corneal oedema
- severe band keratopathy
Complications:
- Transient reduction in acuity following tonometry may occur
Equipment:
- Tonometer
- A simple, rapid and reliable method of measuring IOP
- Dedicated handheld device.
- A magnetic mechanism rapidly presses a small, soft probe against the cornea six times.
- The degree of corneal resistance is translated into an IOP estimate.
- Does not require local anaesthesia or a slit lamp.
- The patient must be sitting upright.
Using the tonometer- ICARE (Video courtesy ICARE USA)
Older eye care tonometer (Video Courtesy ICARE USA)
Procedure:
- Ask the subject to relax and look straight ahead at a specific point.
- Hold the tonometer in a horizontal position.
- Bring the tonometer near to the eye until the probe tip is 4-8 mm from the centre of the cornea.
- The forehead-support adjustment wheel varies this distance.
- Measure IOP by pressing the measurement button.
- After each successful measurement, there is a short beep.
- Perform six measurements.